Instrument parameters | |
light source: | Long life halogen lamp 6V/10W |
Filter wavelength range: | 300nm~800nm |
Test Range: | (-0.3~3.0Abs) |
Resolution: | 0.001Abs (display), 0.0001Abs (internal calculation) |
drift: | ≤0.003Abs/20min |
Colorimetric system: | 10mm micro flow cell capacity 32μL |
Injection volume: | 100μL~9000μL (injection volume accuracy can be automatically compensated) 500μL |
Cuvette temperature: | Peltier temperature control 37℃, 30℃, 25℃, room temperature can be selected arbitrarily, the accuracy is ±0.1℃ |
Cross-contamination rate: | ≤1% |
Repeatability: | CV≤1.0% |
Extreme use temperature: | 5℃~40℃ |
Operating temperature: | 10℃~30℃ |
Relative temperature: | ≤80% |
storage: | Program design can store 320 test items. 20,000 test results |
output: | Standard RS232 bidirectional data interface, parallel printer interface |
power supply: | AC 220×(1±10%)V, (50±)Hz |
Power supply: | 130VA |
Atmospheric pressure: | 75kPa~106kPa |
size: | 392mm×375mm×205mm (length×width×height) |